To start a Disto X calibration, you’ll need to pair it with a personal digital assistant or PDA. The calibration process must be done with an external device that uses calibration measurements downloaded from the Disto X to calculate calibration coefficients. These coefficients are then uploaded back to the Disto X wirelessly via Bluetooth. Note that calibration cannot be run internally on the Disto X.
The first step is to clear any data from the Disto X’s memory. To do this, simultaneously press the ‘auth’ and ‘units’ buttons and hold for two seconds. Afterward, put the Disto X into calibration mode by simultaneously pressing the ‘off’ and ‘smart’ buttons and holding for two seconds. Before starting any calibration, remove anything that might cause interference, such as watches or vertical gear, and move it along with the k-pax away from the calibration area.
To begin collecting the 56 calibration readings, look for two rocks, ledges, or other distinctive features that are roughly 10 feet apart and horizontal. You’ll be taking four measurements in each direction—essentially a set of front sights and back sights. Position the unit on each of its four faces, making sure that it is aligned with the laser shooting along the axis between the two stations.
For the next eight calibration readings, find another pair of horizontal stations oriented roughly 90 degrees from the original stations. Take four front sights and four back sights while rotating around each face of the unit. The remaining 40 measurements are less critical, but shooting between fixed points for each target will generally yield better calibration.
Once you’ve collected all the measurements, reconnect the Disto X to the PDA via Bluetooth and allow the data to download to the calibration software in Pocket Topo. From the calibration menu, press ‘evaluate.’ The software will calculate the delta value—lower delta values indicate better calibration. Generally, a good calibration has a maximum delta of 0.5. If you achieve a low delta, go to the menu and select ‘upload’ to send the new calibration coefficients back to the Disto X.
As a final check on the calibration’s quality, take a front sight and back sight with the unit rotated onto each of its four faces. The total range of measurements when corrected for direction should be no more than about half a degree. If it’s significantly greater, repeat the calibration while ensuring there’s no magnetic interference. Calibrations usually remain valid for up to two months, but it’s a good practice to perform a front sight and back sight check before any new survey.